| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Sep 30 | Class II | SEASPINE ORTHOPEDICS CORPORATION Device |
SeaSpine Mariner Deformity Product Name: Extended Polyaxial Head, Attached Gimbal, 12mm REF: MD1-300010-12 The proximal thread depth of the spinal device is out of design specification. |
| Jul 29 | Class II | SEASPINE ORTHOPEDICS CORPORATION Device |
SeaSpine Polyaxial Head REF Number: 41-3010 Due to manufacturing error, polyaxial head may popoff from the screw intraoperatively following insertion. |
| Dec 24, 2025 | Class II | SEASPINE ORTHOPEDICS CORPORATION Device |
Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF: PC2-400013; Navigation, 3.5mm Drill, REF: PC2-400135; Navigation, 4.0… The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only. |